Skip to main content
. 1986 May-Jun;91(3):113–121. doi: 10.6028/jres.091.018

Table 1.

Performance characteristics of microcalorimetric instrumentation at 25 °C.

Characteristic
calibration constant (F) 16 to 22 W V−1
sensitivity (1/F) 0.045 to 0.059 V W−1
noise level of baseline 20 nV peak-to-peak
detectability (1/3 × noise level/sensitivity) 0.15 μW
stability of baseline 100 nV over 24 hours
linearity of calibration constant 0.1 percent from 0.01 mW to 2 mW
time constanta 63 s
stability of calibration constant 0.2 percent over 2 years
imprecisionb:
 q ≈ 10 mJ 0.3%
 q ≈ 100 mJ 0.1%
 q ≈ 500 mJ 0.05%
 q ≈ 300 mJ (chemical heat) 0.2%
“blank” heat effects:
 no reaction vessel in place  0.023±0.03 mJ
 empty vessel in calorimeter −0.065±0.06 mJ
 H2O  0.67±0.22 mJ
 human serum -0.75±1.0 mJ
glucose isomerase (T = 25 °C) −101±4 mJ
inaccuracyc <0.2%
a

A reaction vessel containing 1.0 mL of water was in the copper container.

b

Based upon experiments in which electrical heat was introduced for periods of 10 to 100 seconds. All uncertainties are standard deviations.

c

The inaccuracy was assessed by the performance of both heater placement tests [1] and by measurements of heats of neutralization of HCl with excess NaOH.