Table 1.
Performance characteristics of microcalorimetric instrumentation at 25 °C.
| Characteristic | |
|---|---|
| calibration constant (F) | 16 to 22 W V−1 |
| sensitivity (1/F) | 0.045 to 0.059 V W−1 |
| noise level of baseline | 20 nV peak-to-peak |
| detectability (1/3 × noise level/sensitivity) | 0.15 μW |
| stability of baseline | 100 nV over 24 hours |
| linearity of calibration constant | 0.1 percent from 0.01 mW to 2 mW |
| time constanta | 63 s |
| stability of calibration constant | 0.2 percent over 2 years |
| imprecisionb: | |
| q ≈ 10 mJ | 0.3% |
| q ≈ 100 mJ | 0.1% |
| q ≈ 500 mJ | 0.05% |
| q ≈ 300 mJ (chemical heat) | 0.2% |
| “blank” heat effects: | |
| no reaction vessel in place | 0.023±0.03 mJ |
| empty vessel in calorimeter | −0.065±0.06 mJ |
| H2O | 0.67±0.22 mJ |
| human serum | -0.75±1.0 mJ |
| glucose isomerase (T = 25 °C) | −101±4 mJ |
| inaccuracyc | <0.2% |
A reaction vessel containing 1.0 mL of water was in the copper container.
Based upon experiments in which electrical heat was introduced for periods of 10 to 100 seconds. All uncertainties are standard deviations.
The inaccuracy was assessed by the performance of both heater placement tests [1] and by measurements of heats of neutralization of HCl with excess NaOH.