Table 2. Calibrator reactivity and reproducibility of results assessed by repeat testing.

Specimen Maxim Sedia
N Mean OD CV OD (%) Mean ODn CV ODn (%) N Mean OD CV OD (%) Mean ODn CV ODn (%)
Calibrators
Alla 222 0.75 10.4 N/A N/A 219 0.65 15.0 N/A N/A
Median valuesb 74 0.75 9.3 N/A N/A 73 0.64 14.2 N/A N/A
Kit-supplied control specimens
Acute (low) 222 0.36 10.5 0.49 8.0 219 0.35 16.2 0.55 14.0
Chronic (high) 222 1.37 8.6 1.83 7.9 219 1.31 10.4 2.06 12.7
Blinded control specimens
BC-1 25 3.30 5.1 4.45 9.3 25 3.07 5.6 4.94 15.0
BC-2 25 3.04 5.9 4.08 8.9 25 2.83 6.2 4.5 13.2
BC-3 25 0.40 14.2 0.54 14.8 25 0.67 20.0 1.02 13.6

aAverage over all calibrator values;

bAverage over median calibrator values (one value per plate).