Skip to main content
. 2019 Feb 6;5:3. doi: 10.1186/s41205-019-0040-4

Table 1.

Fitting Parameters for Calibration Curve

Color Channel a b c
Red − 582.230 − 436.180 0.757
Green − 996.825 − 727.023 0.730
Blue − 1633.021 − 773.091 0.475

A rational function, y = a + (b/(x – c)), was used to fit the data where y is the dose and x is the NOD. Based on these parameters, a calibration curve can be created for each color channel