Table 3.

Summary of statistical parameters needed to assess the trueness of the approach, based on measurement of the modal equivalent circular diameter (ECD) of ERM-FD100 and ERM-FD304 at magnifications of 18,500× and 68,000×.

Material Magnification Cm (nm) uc(x) (nm) CCRM (nm) uCRM (nm) Δm (nm) uΔ (nm) UΔ (nm)
ERM-FD100 18,500× 19.2 0.9 19.4 0.7 0.2 1.1 2.3
68,000× 19.4 0.8 19.4 0.7 0.0 1.1 2.1
ERM-FD304 18,500× 25.5 1.3 27.8 0.8 2.3 1.5 3.1
68,000× 24.9 1.0 27.8 0.8 2.9 1.3 2.6