Table 2.

Description of mp-MRI parameters

Sequence Coil TR TE FA degrees WFS (pix) BW Hz/Px FoV (mm) SL (mm) Gap TSE factor PD FS ACQ
matrix
TRs (s) Total scan
duration
T2 TSE coronal Dual 6128 100 90 2.704 160.7 180 3 3 16 R > L No 300 × 290 05:55.4
T2 TSE axial Dual 5407 100 90 2.704 160.7 180 3 0 16 R > L No 300 × 290 05:13.6
T1W TSE Dual 487 8 90 1.997 217.6 240 3 3 4 R > L No 184 × 184 03:06.8
DWI 01505001000 Dual 2753 80 90 40.353 10.8 220 5 0 A > P SPAIR 168 × 169 05:16.5
DWI b2000 Dual 2000 78 90 44.108 9.9 220 5 0 A > P SPIR 168 × 169 03:40.0
DCE Dual 5.8 2.8 10 1.766 246.1 180 3 0 R > L SPAIR 140 × 162 13 04:14.1

TSE, turbo spin echo; TR, time to repetition; TE, time to echo; FA, flip angle; WFS, water-fat shift; BW, bandwidth; FoV, field of view; DWI, diffusion-weighted imaging; DCE, dynamic contrast-enhanced; TRs, temporal resolution; PD, phasing direction; SL, slice thickness