Table 2.

Attenuation coefficient inter-platform reproducibility for measurements based on a single acquisition (N=1) and on the mean of multiple repeated intra-platform acquisitions (N=2, 3, 4, and 5).

Number of
acquisitions N
GE mean ± SD
(dB/cm-MHz)
Siemens mean ±
SD (dB/cm-MHz)
ICC (95% CI) wCV
1 0.89 ± 0.13 0.94 ± 0.15 0.69 (0.50, 0.81) 0.09
2 0.89 ± 0.13 0.94 ± 0.15 0.76 (0.58, 0.86) 0.08
3 0.89 ± 0.13 0.94 ± 0.15 0.75 (0.57, 0.86) 0.08
4 0.90 ± 0.13 0.94 ± 0.15 0.77 (0.58, 0.86) 0.07
5 0.90 ± 0.13 0.94 ± 0.15 0.77 (0.59, 0.87) 0.07

ICC was calculated based on a single-unit, absolute-agreement, two-way mixed effect analysis of variance (ANOVA) model.

CI confidence interval, ICC intraclass correlation coefficient, SD standard deviation, wCV within-subject coefficient of variation