Table 3.

Log-transformed backscatter coefficient inter-platform reproducibility for measurements based on a single acquisition (N=1) and on the mean of multiple repeated intra-platform acquisitions (N=2, 3, 4, and 5).

Number of
acquisitions N
GE mean ±
SD (dB)
Siemens mean ±
SD (dB)
ICC (95% CI)
1 −30.8 ± 4.9 −27.8 ± 5.7 0.67 (0.19, 0.85)
2 −30.7 ± 4.9 −27.7 ± 5.7 0.68 (0.23, 0.85)
3 −30.7 ± 5.0 −27.8 ± 5.7 0.69 (0.24, 0.85)
4 −30.6 ± 5.0 −27.8 ± 5.7 0.70 (0.26, 0.86)
5 −30.6 ± 5.0 −27.9 ± 5.6 0.70 (0.28, 0.86)

ICC was calculated based on a single-unit, absolute-agreement, two-way mixed effect analysis of variance (ANOVA) model.

CI confidence interval, ICC intraclass correlation coefficient, SD standard deviation