Skip to main content
. 2019 Aug 1;32(4):224–241. doi: 10.1089/jamp.2018.1488

Table 5.

Experimentally Measured Mean (Standard Deviation) Temperature and Relative Humidity Values for All Mixer Heater Designs with a Targeted 32°C Outlet Temperature Operating in Alternating Mode

  20 LPM 30 LPM
Design 1
 T (C) 32.7 (0.5) 32.2 (0.6)
 RH (%) 33 (2.1) 32 (1.7)
Design 2
 T (C) 32.8 (0.5) 32.0 (0.7)
 RH (%) 39 (1.1) 30 (0.0)
Design 3
 T (C) 32.7 (0.3) 32.4 (0.5)
 RH (%) 35 (0.6) 30 (1.1)