Table 1.

List of exposure parameters and reconstruction techniques used in the reference and ultra-low-dose test protocols

Exposure protocol Reconstruction technique Matrix In-plane voxel size (mm) kV mA Rotation time (s) Pitch CTDIvol (mGy) Effective dose (µSv)a
Scan length of 5 cm Scan length of 10 cm
Reference FBP 512 × 512 0.391 120 100 1.0 0.53 29.4 308.7 617.4
LD1 FBP 512 × 512 0.391 80 10 0.4 0.53 0.53 5.6 11.1
ASIR 50
ASIR 100
MBIR 1024 × 1024 0.195
LD2 FBP 512 × 512 0.391 0.97 0.29 3.0 6.1
ASIR 50
ASIR 100
MBIR 1024 × 1024 0.195

CTDIvol volume CT dose index, LD low-dose protocol, FBP filtered back-projection, ASIR adaptive statistical iterative reconstruction, MBIR model-based iterative reconstruction

aCalculated from the dose–length product based upon a k factor of 0.0021