Table 1.
List of exposure parameters and reconstruction techniques used in the reference and ultra-low-dose test protocols
| Exposure protocol | Reconstruction technique | Matrix | In-plane voxel size (mm) | kV | mA | Rotation time (s) | Pitch | CTDIvol (mGy) | Effective dose (µSv)a | |
|---|---|---|---|---|---|---|---|---|---|---|
| Scan length of 5 cm | Scan length of 10 cm | |||||||||
| Reference | FBP | 512 × 512 | 0.391 | 120 | 100 | 1.0 | 0.53 | 29.4 | 308.7 | 617.4 |
| LD1 | FBP | 512 × 512 | 0.391 | 80 | 10 | 0.4 | 0.53 | 0.53 | 5.6 | 11.1 |
| ASIR 50 | ||||||||||
| ASIR 100 | ||||||||||
| MBIR | 1024 × 1024 | 0.195 | ||||||||
| LD2 | FBP | 512 × 512 | 0.391 | 0.97 | 0.29 | 3.0 | 6.1 | |||
| ASIR 50 | ||||||||||
| ASIR 100 | ||||||||||
| MBIR | 1024 × 1024 | 0.195 | ||||||||
CTDIvol volume CT dose index, LD low-dose protocol, FBP filtered back-projection, ASIR adaptive statistical iterative reconstruction, MBIR model-based iterative reconstruction
aCalculated from the dose–length product based upon a k factor of 0.0021