Table 4.

Differences between linear measurements obtained by the reference protocol (reference dose/filtered back-projection) and the various test protocols

Mean difference (mm)* Standard deviation of difference (mm) 95% confidence interval limits (mm) Significance (linear regression)
Lower limit Upper limit
LD1/FBP − 0.2083 0.61128 − 0.3662 − 0.0504 0.605
LD1/ASIR 50 − 0.2817 0.66498 − 0.4534 − 0.1099 0.394
LD1/ASIR 100 − 0.3467 0.61959 − 0.5067 − 0.1866 –
LD1/MBIR − 0.4667 1.01508 − 0.7289 − 0.2044 –
LD2/FBP − 0.2500 0.76878 − 0.4486 − 0.0514 0.584
LD2/ASIR 50 − 0.6367 1.47269 − 1.0171 − 0.2562 –
LD2/ASIR 100 − 0.4050 0.55736 − 0.5490 − 0.2610 –
LD2/MBIR − 0.4583 1.05899 − 0.7319 − 0.1848 –

– Bland–Altman plot and linear regression were not performed because the one-sample t test demonstrated a mean difference greater than ± 0.3 mm with 95% confidence limits greater than ± 0.5 mm

LD low-dose protocol, FBP filtered back-projection, ASIR adaptive statistical iterative reconstruction, MBIR model-based iterative reconstruction

*All mean differences were statistically significant (p < 0.05)