Table 1.

Reference Acquisition Parameters for Contrast Enhanced Cardiovascular Magnetic Resonance (CE-MRA), 3D whole-heart Balanced State Free Precession (bSSFP) and modified Dixon (mDixon) sequences

Parameter CE-CMRA bSSFP mDixon
Average Acquisition Time Goal Dynamic scan time goal ~ 12 s 3:11 min 3:08 min
TE (ms) 0.87 2

1st image: 1.8

2nd image: 3.4

TR (ms) 2.2 4.5 5.2
Flip Angle (degrees) 30° 90° 15°
FOV (mm) 260 × 260 260 × 260 260 × 260
Acquired Voxel Size 1.4 × 1.4 × 2.8 1.4 × 1.4 × 1 1.4 × 1.4 × 1
Recon Voxel (mm3) 0.9 × 0.9 × 0.4 0.9 × 0.9 × 1 0.9 × 0.9 × 1
Sense LR: 2.5; AP: 1.0 LR: 1.5; AP: 1.0 LR: 2.0; AP: 1.5
Trigger Delay NA Mid-Diastole Mid-Diastole
Shot Duration (ms) NA 130 130
Navigator Window (mm) Cartesian k-space acquisition window 7 7
Slice number 90 135 135
Bandwidth (Hz/pixel units) 0.165 0.35 0.3

Abbreviations: mDixon modified-Dixon, bSSFP balanced Steady State Free Precession, CE-CMRA Contrast-Enhanced Cardiovascular Magnetic Resonance Angiography; TE echo time; TR repetition time; NA not applicable; LR Left Right; AP Anterior Posterior