Skip to main content
. 2019 Aug 7;45:97. doi: 10.1186/s13052-019-0686-y

Table 1.

Differences between Type C and D Ambulance

List of main differences between Type C and Type D ground ambulance
Equipment EU EN 1789 Type C ground ambulance Our suggested Type D neonatal transport ground ambulance
Mechanical ventilator Adult use only Neonatal use only
Oxygen cylinders Present Present
Air compressed cylinders Not planned Needed
iNO cylinders and delivery system Not planned Needed
Phototheraphy Not planned Needed
Shock adsorbing stretcher support Not planned Needed
Second mechanical ventilator (twin newborn transport) Not planned Needed (mounted on transport module)
Defibrillator Adult use only Neonatal use only
Cooling device Not planned Needed
Patient’s placenta (refrigeration box) Not planned Needed
Transcutaneous gas analyzer Not planned Needed (mounted on transport module)