Table 1.
Statistical analysis for PBF and DGF of various syringe and needle combinations
| PBF (N) | DGF (N) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|||||||||
| Mean difference | 95%CI | P-valueb | F-statistic (df1,df2)a;P-valuea | Mean difference | 95%CI | P-valueb | F-statistic (df1,df2)a;P-valuea | |||
| 1 mL syringe | ||||||||||
| ORI | versus | 27GN | 0.28 | −0.19, 0.75 | 0.420 | F(3, 8) = 3.545; P = 0.068 | −0.19 | −0.30, −0.07* | 0.003 | F(3, 8) = 103.881; P = 0.000 |
| ORI | versus | 27GSN | 0.40 | −0.07, 0.86 | 0.109 | −0.23 | −0.35, −0.12* | 0.001 | ||
| ORI | versus | 30GN | 0.35 | −0.11, 0.82 | 0.179 | −0.58 | −0.70, −0.46* | 0.000 | ||
| 27GN | versus | 27GSN | 0.12 | −0.35, 0.58 | 1.000 | −0.05 | −0.16, 0.07 | 1.000 | ||
| 27GN | versus | 30GN | 0.07 | −0.39, 0.54 | 1.000 | −0.39 | −0.51, −0.28* | 0.000 | ||
| 27GSN | versus | 30GN | −0.04 | −0.51, 0.42 | 1.000 | −0.35 | −0.46, −0.23* | 0.000 | ||
| 3 mL syringe | ||||||||||
| ORI | versus | 27GN | −0.65 | −1.80, 0.50 | 0.510 | F(3, 8) = 34.845; P < 0.001 | −1.91 | −2.59, −1.23* | 0.000 | F(3, 8) = 1051.939; P < 0.001 |
| ORI | versus | 27GSN | −2.42 | −3.57, −1.27* | 0.001 | −3.42 | −4.11, −2.74* | 0.000 | ||
| ORI | versus | 30GN | −2.88 | −4.04, −1.73* | 0.000 | −10.35 | −11.03, −9.66* | 0.000 | ||
| 27GN | versus | 27GSN | −1.77 | −2.92, −0.62* | 0.004 | −1.51 | −2.20, −0.83* | 0.000 | ||
| 27GN | versus | 30GN | −2.23 | −3.38, −1.08* | 0.001 | −8.44 | −9.12, −7.75* | 0.000 | ||
| 27GSN | versus | 30GN | −0.47 | −1.62, 0.69 | 1.000 | −6.93 | −7.61, −6.24* | 0.000 | ||
| 5 mL syringe | ||||||||||
| ORI | versus | 27GN | −8.09 | −10.22, −5.96* | 0.000 | F(3, 8) = 108.173; P < 0.001 | −11.67 | −12.33, −11.01* | 0.000 | F(3, 8)= 33503.958; P < 0.001 |
| ORI | versus | 27GSN | −8.23 | −10.36, −6.10* | 0.000 | −17.67 | −18.33, −17.00* | 0.000 | ||
| ORI | versus | 30GN | −10.14 | −12.27, −8.01* | 0.000 | −56.70 | −57.36, −56.04* | 0.000 | ||
| 27GN | versus | 27GSN | −0.14 | −2.27, 1.99 | 1.000 | −6.00 | −6.66, −5.34* | 0.000 | ||
| 27GN | versus | 30GN | −2.05 | −4.18, 0.08 | 0.061 | −45.03 | −45.69, −44.37* | 0.000 | ||
| 27GSN | versus | 30GN | −1.91 | −4.04, 0.22 | 0.086 | −39.03 | −39.69, −38.37* | 0.000 | ||
| 10 mL syringe | ||||||||||
| ORI | versus | 27GN | −1.47 | −30.11, 27.18 | 1.000 | F(3, 8)= 12.992; P = 0.002 | −34.36 | −39.24, −29.48* | 0.000 | F(3, 8)= 6114.772; P < 0.001 |
| ORI | versus | 27GSN | −1.07 | −29.71, 27.58 | 1.000 | −41.01 | −45.88, −36.13* | 0.000 | ||
| ORI | versus | 30GN | −42.80 | −71.45, −14.15* | 0.005 | −175.95 | −180,83, −171.08* | 0.000 | ||
| 27GN | versus | 27GSN | 0.40 | −28.25, 29.15 | 1.000 | −6.65 | −11,53, −1.77* | 0.009 | ||
| 27GN | versus | 30GN | −41.33 | −70.00, −12.69* | 0.006 | −141.60 | −146.47, −136.72* | 0.000 | ||
| 27GSN | versus | 30GN | −41.73 | −70.38, −13.09* | 0.006 | −134.95 | −139.82, −130.07* | 0.000 | ||
| 27GN | versus | 27GSN | 3.00 | 0.58, 5.42* | 0.020 | F(2, 6) = 766.569; P < 0.001 | 21.27 | 18.68, 23.86* | 0.000 | F(2, 6) = 16666.452; P < 0.001 |
P < 0.05,
One-way ANOVA,
Post-hoc analysis with Bonferroni corrections
DGF (dynamic gliding force), N (Newtons), ORI (original needle from packaging of the syringe), PBF (plunger-stopper break-loose force), df1 (the numerator degree of freedom), df2 (the denominator degree of freedom), 27GN (27 Gauge needle), 27GSN (27 Gauge spinal needle), 30GN (30 Gauge needle)