Skip to main content
. 2019 Aug 12;9:11618. doi: 10.1038/s41598-019-48039-x

Figure 3.

Figure 3

Nominal vs. measured R using calibrated resistors of 0.19, 0.49 and 0.68 kPa.s.L−1. Dashed line represents the theoretical line of identity if a device correctly captured the resistance value. The slopes of the tremoflo vs. IOS are significantly different (p = 0.003), the IOS slope is also significantly different from the theoretical line of identity (p = 0.0028), whereas the tremoflo slope is not significantly different from the theoretical line of identity (p = 0.057).