Table 2.

Results of measurements with oscillometry.

Variable Instrument All subjects COPD Healthy subjects
R5 (kPa.s.L−1) IOS 0.36 ± 0.09 0.39 ± 0.08 0.31 ± 0.08***
tremoflo 0.35 ± 0.12 0.41 ± 0.11 0.26 ± 0.07
R20 (kPa.s.L−1) IOS 0.29 ± 0.07 0.30 ± 0.08 0.40 ± 0.05***
R19 (kPa.s.L−1) tremoflo 0.29 ± 0.06 0.31 ± 0.06 0.25 ± 0.04
R5-R20 (kPa.s.L−1) IOS 0.05 ± 0.06 0.08 ± 0.06 0.02 ± 0.05
R5-R19 (kPa.s.L−1) tremoflo 0.06 ± 0.09 0.10 ± 0.09 0.00 ± 0.04
X5 (kPa.s.L−1) IOS −0.13 ± 0.08*** −0.15 ± 0.09*** −0.09 ± 0.03
tremoflo −0.20 ± 0.17 −0.27 ± 0.18 −0.10 ± 0.04
fres (Hz) IOS 12.5 ± 4.4*** 14.4 ± 4.4*** 9.4 ± 2.3***
tremoflo 17.0 ± 6.7 20.8 ± 5.6 10.9 ± 2.6**
AX (kPa.s.L−1) IOS 0.55 ± 0.59*** 0.75 ± 0.66*** 0.21 ± 0.17
tremoflo 1.34 ± 1.62 1.99 ± 1.78 0.31 ± 0.22

Mean ± SD. Significant differences between IOS and tremoflo: **p < 0.01, ***p < 0.001, paired, two-tailed t-test.