. 2019 Jun 23;46(8):3700–3708. doi: 10.1002/mp.13638

Table 2.

Sensitivity and specificity of the Mobius3D recalculation as compared to institutional IMRT QA for failing phantom plans. The independent recalculation was consistently more sensitive to detecting unacceptable plans.

Device # tests # failing results % Sensitivity % Specificity
IMRT QA Mobius3D Sig. IMRT QA Mobius3D Sig.
All 337 18 6 72 ** 98 68 **
EPID 58 7 0 71 * 100 57 **
ArcCheck 93 4 0 100 * 100 70 **
Ion Chamber 44 1 0 100   91 67 *
IC + Array 29 0 N/A N/A   93 62 *
MapCheck 121 5 20 40   100 67 **
*

Significant (0.001–0.05);

**

Highly significant (< 0.001).