Table 3.

Sensitivity and specificity of the Mobius3D recalculation as compared to institutional IMRT QA for poor phantom results (>5% dose deviations). The independent recalculation was consistently more sensitive to detecting unacceptable plans.

Device # tests # poor results % Sensitivity % Specificity
IMRT QA Mobius3D Sig. IMRT QA Mobius3D Sig.
All 337 59 5 47 ** 99 69 **
EPID 58 16 0 56 * 100 57 **
ArcCheck 93 16 0 31 * 100 66 **
Ion Chamber 44 8 25 63   94 72 *
IC + Array 29 6 17 50   96 65 *
MapCheck 121 20 5 45 * 100 69 **
*

Significant (0.001–0.05);

**

Highly significant (< 0.001).