Table A1a.

Resistance of thermometer 80179 at fixed points.

Run 1 Run 2

FP1 R(TP)2 R (FF)3 HE(4)4 R(TP) R(FP) HE(4)
TP 0,24354530 10 0.24355153 23
AU 1.11338142 17 1.11339495 30
TP 0.24354484 37 0.24355032 19
AG 1.04390980 22 1.04391980 11
TP 0.24354602 10 0.24355100 4
AL 0.82218768 15 0.82219719 20
TP 0.24354487 19 0.24355033 1
ZN 0.62563109 15 0.62563938 9
TP 0.24354467 22 0.24355013 8
SN 0.46097202 24 0.46097844 28
TP 0.24354465 16 0.24354968 14
1

Measurements taken in order indicated.

2

R (TP): thermometer resistance (ohms) at triple point of water for zero measuring current.

3

R(FP): thermometer resistance (ohms) at indicated metal freezing point for zero measuring current.

4

HE(4): increase in resistance of thermometer (ohms × 10−8) due to heating effect of 4 mA measuring current.