Table A3a.

Immersion characteristics of thermometers in zinc cell. 1,3

Station2 Resistance of thermometer number Station Resistance of thermometer number

cm 80179 18227 18236 18237 cm 80179 18227 18236 18237
0 0.62564094 0.66881727 0.66111506 0.67483970 7 0.62563817 0.66881563 0.66111550 0.67483869
0 0.62564009 0.66881635 0.66111560 0.67483926 6 0.62563851 0.66881562 0.66111561 0.67483881
0 0.62563969 0.66881612 0.66111562 0.67483920 0 0.62563869 0.66881598 0.66111585 0.67483930
0 0.62563923 0.66881629 0.66111546 0.67483921 0 0.62563858 0.66881588 0.66111580 0.67483918
18 0.62469878 0.66772854 0.65999691 0.67377628 0 0.62563850 0.66881599 0.66111593 0.67483931
18 0.62468420 0.66769737 0.65997147 0.67371738 5 0.62563830 0.66881572 0.66111552 0.67483899
16 0.62532066 0.66845569 0.66074723 0.67446863 4 0.62563851 0.66881563 0.66111576 0.67483919
14 0.62555932 0.66871951 0.66101985 0.67474217 3 0.62563848 0.66881566 0.66111565 0.67483923
12 0.62562814 0.66880302 0.66110364 0.67482532 2 0.62563854 0.66881561 0.66111586 0.67483919
10 0.62563737 0.66881433 0.66111437 0.67483735 1 0.62563843 0.66881586 0.66111592 0.67483946
0 0.62563886 0.66881583 0.66111589 0.67483939 0 0.62563836 0.66881582 0.66111589 0.67483920
0 0.62563878 0.66881593 0.66111572 0.67483913 0 0.62563841 0.66881591 0.66111581 0.67483919
0 0.62563872 0.66881580 0.66111571 0.67483926 0 0.62563846 0.66881569 0.66111578 0.67483930
10 0.62563703 0.66881444 0.66111416 0.67483761 0 0.62563843 0.66881570 0.66111577 0.67483935
9 0.62563784 0.66881497 0.66111503 0.67483833 0 0.62563840 0.66881587 0.66111591 0.67483913
8 0.62563799 0.66881528 0.66111545 0.67483868
1

Resistance determinations made at 5 min intervals.

2

Station: distance (cm) thermometer raised above full immersion.

3