. 1984 Sep-Oct;89(5):349–373. doi: 10.6028/jres.089.020

Table A2a.

Effect of guarding at silver point, thermometer 80179.

No.1 Thermometer resistance2 No. Thermometer resistance

External Guard No Guard External Guard No Guard
1 1.04392444 31 1.04392563
2 1.04392494 32 1.04392576
3 1.04392533 33 1.04392535
4 1.04392506 34 1.04392155
5 1.04392526 35 1.04392162
6 1.04392090 36 1.04392582
7 1.04392092 37 1.04392580
8 1.04392524 38 1.04392595
9 1.04392531 39 1.04392578
10 1.04392552 40 1.04392577
11 1.04392542 41 1.04392101
12 1.04392569 42 1.04392163
13 1.04392100 43 1.04392580
14 1.04392127 44 1.04392567
15 1.04392567 45 1.04392564
16 1.04392553 46 1.04392575
17 1.04392556 47 1.04392592
18 1.04392541 48 1.04392127
19 1.04392533 49 1.04392115
20 1.04392126 50 1.04392502
21 1.04392155 51 1.04392489
22 1.04392592 52 1.04392467
23 1.04392583 53 1.04392454
24 1.04392610 54 1.04392402
25 1.04392569 55 1.04391980
26 1.04392554 56 1.04391904
27 1.04392194 57 1.04392274
28 1.04392176 58 1.04392240
29 1.04392554 59 1.04392180
30 1.04392560 60 1.04392091
1

No: sequence number of reading. Resistance determinations made at 5 min intervals.

2

All determinations made with normal measuring current.