Skip to main content
. Author manuscript; available in PMC: 2020 Sep 1.
Published in final edited form as: IEEE Trans Instrum Meas. 2018 Nov 15;68(9):3200–3210. doi: 10.1109/TIM.2018.2877952

Fig. 11.

Fig. 11.

Results of the SiPM V-I curve experiments, (a) Log-linear plot of the V-I curves for 600 SiPMs measured with the 100-channel Pico-PET system, (b) Distribution of the measured breakdown voltages of the 600 SiPMs (mean ± STD: 24.6 V ± 0.089 V). (c) Linear scale plot of the 600 V-I curves. The typical and maximum dark currents provided by the SiPM vendor are marked in the plot with horizontal lines.