Skip to main content
. 2019 Aug 15;9:11892. doi: 10.1038/s41598-019-48386-9

Figure 7.

Figure 7

Agreement between radar and ECG in the PeAF group. The PR (A), average R-R interval (B) and individual R-R intervals (C) were measured in the PeAF group using IR-UWB radar and ECG. Both the PR and average R-R interval showed strong agreement between radar and ECG. The level of agreement for the individual R-R intervals between the two methods was significantly lower than that for the PR and average R-R intervals but was still sufficiently high. ECG, electrocardiography; ICCR, intraclass correlation coefficient.