Skip to main content
. 2019 Aug 1;20(15):3761. doi: 10.3390/ijms20153761

Table 1.

Quantitative analysis of micro-CT imaging.

Follow-Up Period (Weeks) 6 12
PEO-Coated Group Non-Coated Group p Value ** PEO-Coated Group Non-Coated Group p Value **
No. of animals 6 6 6 6
Residual volume
(RV) (mm3)
42.5 (40.1 to 44.5) * 33.6 (30.9 to 37.6) * 0.010 36.1 (30.3 to 40.6) * 28.4 (23.5 to 33.2) * 0.026
Void volume
(VV) (mm3)
2.4 (1.1 to 4.3) * 19.4 (13.9 to 34.2) * 0.010 19.4 (8.6 to 33.7) * 42.1 (15.5 to 67.3) * 0.132
Degradation percentage (%) 5.3 (0.8 to 10.6) * 25.2 (31.2 to 16.2) * 0.010 19.6 (0.6 to 32.5) * 36.7 (26.1 to 47.5) * 0.026
Degradation rate (mm/year) 0.145
(0.021 to 0.292) *
0.690
(0.445 to 0.856) *
0.010 0.269
(0.131 to 0.446) *
0.503
(0.357 to 0.651) *
0.026

* Data in parenthesis represent ranges; ** Mann-Whitney test; Degradation rate was calculated as follows [11]: degradation rate (mm/year) = initial volume  RV  initial surface area x time; where initial volume is 44.87 mm3 and the initial surface area is 141.83 mm2.