Table 2.

Mean values of artefact percentages and signal-to-noise ratios (SNR) for measurements using gel electrodes and the electrocardiography (ECG)-belt in clinical (data from Fontana et al. [9]) and home settings (n = 12).

Gel Electrode, Clinics ECG-Belt, Clinics ECG-Belt, Home
Artefact percentages
 Mean % 2.9 9.7 7.5
 Median % 1.4 5.4 2.8
 SD % 4.1 14.7 10.8
 Inter-patient CV % 144.4 151.7 143.6
 Intra-patient CV % - - 2.8
SNR
 SNRlf dB 12 0 11
 SD dB 5 5 2
 BLW mV 0.03 0.30 0.05
 SD mV 0.02 0.43 0.02
 SNRhf dB 21 17 21
 SD dB 3 6 5

SD—standard deviation; CV—coefficient of variation; Intra-patient CV was calculated as night-to-night variation; SNRlf—low frequency SNR; BLW—baseline wander; SNRhf—high frequency SNR.