Table 1.

B0 homogeneity. Standard deviation values of the frequency, measured with and without the TxS system for tune-up shim and automatic shim. Data were evaluated either for the entire phantom or only for the slice shown in Figure 6.

Standard Deviation of ∆f (Hz)
with TxS System without TxS System
tune-up shim phantom 97.0 97.5
slice 40.4 36.8
automatic shim phantom 72.5 75.9
slice 45.4 43.0