Table 3.
Comparison of assay sensitivities between 5-plex and duplex reactions on clinical positive samples
| IAV clinical sample | ICV clinical sample | IDV clinical sample | |||||
|---|---|---|---|---|---|---|---|
| Duplex | 5-plex | Duplex | 5-plex | Duplex | 5-plex | ||
| Correlation coefficient R2 | 0.998 | 0.993 | 0.997 | 0.990 | 0.996 | 0.996 | |
| PCR efficiency (E) | 90.12% | 90.01% | 95.22% | 94.73% | 90.69% | 99.58% | |
| Ct on serial dilutions* | 10−1 | 21.03 | 21.00 | 22.61 | 22.62 | 20.52 | 20.43 |
| 10−2 | 24.40 | 24.24 | 26.06 | 26.09 | 23.53 | 23.66 | |
| 10−3 | 27.91 | 27.73 | 29.61 | 29.52 | 26.95 | 26.95 | |
| 10−4 | 31.66 | 31.20 | 33.19 | 33.19 | 31.04 | 30.31 | |
| 10−5 | 35.14 | 34.55 | 36.25 | 36.03 | 34.34 | 33.45 | |
| 1/2*10−5 | 36.30 | 35.90 | 37.52 | 37.35 | 35.54 | 34.31 | |
| 1/4*10−5 | 37.42 | 37.17 | 38.46 | 38.87 | 36.24 | 35.57 | |
| 1/8*10−5 | 38.52 | 39.07 | N/A | N/A | 38.1 | 37.2 | |
| Ct difference between assays | 0.33 | 0.13 | 0.58 | ||||