Table 1.

Physics plan check elements with the highest severity (>7).

Physics plan check items Severity
Rx v plan: site 9.3
Rx v plan: laterality 9.1
Contouring: target(s) 8.4
SIM: consistency between orientation of image on the CT scan and treatment plan 8.4
Rx v plan: total dose 8.4
Rx v plan: dose /fraction 8.3
Rx v plan: number of fractions 8.3
Patient assessments: cardiac device, fetus, etc. are taken into consideration for RT 8.0
Isocenter shift: user origin is set correctly 8.0
Isocenter shift: journal note with correct shifts and table top 7.9
Plan: MU (e.g., correct MU for 2D plan) 7.9
Patient assessments: previous RT taken into consideration 7.6
Plan: collision 7.3
Contouring: organs‐at‐risk (OAR's) 7.1

Abbreviations: Rx – prescription; SIM – simulation; RT – radiotherapy; MU – monitor unit.