Skip to main content
. 2019 Aug 21;9:12223. doi: 10.1038/s41598-019-48685-1

Table 3.

Bland-Altman analysis and ICC of pairwise comparison to assess intraobserver agreement for LV volumes, mass, ejection fraction and strain parameters (analysis of entire study population, n = 30).

Parameter First
measurement
Second
measurement
Mean
difference
Limits of
agreement
ICC
(95% CI)
LVEDVi (ml/m2) 97.84 ± 29.26 99.36 ± 30.31 −1.52 ± 9.42 −19.98 to 16.95 0.98 (0.95 to 0.99)
LVESVi (ml/m2) 56.05 ± 34.16 60.38 ± 35.90 −4.33 ± 8.13 −20.27 to 11.60 0.98 (0.95 to 0.99)
LVMi (g/m2) 70.19 ± 20.20 64.45 ± 16.45 5.75 ± 9.34 −12.57 to 24.06 0.91 (0.74 to 0.96)
LVEF (%) 46.58 ± 17.02 43.10 ± 17.01 3.48 ± 4.76 −5.85 to 12.81 0.97 (0.88 to 0.99)
GLS (%) −16.74 ± 5.03 −16.60 ± 4.87 −0.14 ± 9.11 −1.93 to 1.65 0.99 (0.98 to 1.00)
GCS (%) −16.61 ± 4.47 −16.28 ± 4.58 −0.33 ± 0.95 −2.19 to 1.54 0.99 (0.97 to 0.99)

Results are reported as mean ± standard deviation. Abbreviations as in Table 1.