Table 1.
Step | Problem | Possible reason/cause | Solution/suggestions |
4.5 | Low concentration of RNA in the homogenate | • The pellet is not completely resuspended | • Resuspend again by pipetting up and down and incubate on ice for 10–15 min |
5.5 | Low polyribosome peaks are observed | • Degradation of RNA • “Runoff” of ribosomes |
• Increase the quantity/concentration of RNAse inhibitors • Keep the temperature low (0–4); Keep all the solutions and materials cold and RNAse free • Tumors must be cooled/frozen as quickly as possible after harvesting and stored at –80; Keep all solutions cold and work on ice |
5.5 | Nothing is detected in the polyribosome profiles | • Polyribosome extraction didn’t work properly • Poor quality of the sample • Not enough tumor tissue material was used • Sensitivity is too low |
• Increase the number of strokes during the mechanical lysis (add one or 2 more strokes) • Tumors must be cooled/frozen as quickly as possible after harvested and stored at –80 • For optimal polyribosomal profiles results, it is recommended to load a 10–20 OD260 • Increase the sensitivity of the UA-6 absorbance monitor by changing the setting to a higher sensitivity |
5.5 | Poor quality profiles | • Disrupted gradients | • The gradients should be handled with caution in order to minimize any perturbation |