Table 4.
Architect (threshold) | Sensitivity (%) | Specificity (%) | Platelia (threshold) | Sensitivity (%) | Specificity (%) | Elecsys (threshold) | Sensitivity (%) | Specificity (%) |
---|---|---|---|---|---|---|---|---|
0.00 | 100.00 | 0.00 | 0.00 | 100.00 | 0.00 | 0.13 | 100.00 | 0.00 |
0.30a | 97.56 | 85.29 | 0.53 | 98.78 | 79.78 | 0.47 | 100.00 | 85.06 |
0.40 | 93.90 | 92.28 | 1.02a | 95.12 | 88.24 | 1.12c | 100.00 | 85.66 |
0.50 | 91.46 | 94.85 | 1.58 | 91.46 | 94.85 | 6.70a | 100.00 | 93.38 |
0.90 | 74.39 | 98.53 | 2.43 | 78.05 | 98.16 | 9.43 | 98.78 | 95.22 |
1.00 | 69.51 | 98.90 | 4.76b | 53.66 | 100.00 | 20.51b | 90.24 | 99.26 |
1.40b | 54.88 | 100.00 | 5.90c | 37.80 | 100.00 | 30.13c | 73.17 | 99.26 |
1.60c | 43.90 | 100.00 | 9.09c | 32.93 | 100.00 | 30.57 | 71.95 | 100.00 |
New threshold for grey zone.
New threshold for positive zone.
Values close to the supplier thresholds.