2 x 2 Contingency table and statistical analysis assessing the accuracy of RedLine Alert Assay based on the data from Phase 1 testing and all clinical panel microorganisms in Phase 2

Result Anthrax Positive Anthrax Negative Total
Red Line Alert Test positive 64 0 64
Red Line Alert Test negative 4 320 324
Total 68 320 388
Parameter Percentage Confidence Interval
Sensitivity 100.00% 94.40% to 100.00%
Specificity 98.77% 96.87% to 99.66%
Accuracy 98.97% 97.38% to 99.72%
Area under the curve 0.99 (0.99 to 1.00)
Negative predictive value 100.00%