Skip to main content
. 2019 Aug 19;21(9):522–530. doi: 10.1089/dia.2019.0148

Table 4.

Results from Comparative Study

Comparative performance measures GAI GEK
Successful dose administrations 14/16 (88%) 5/16 (31%)
Administered with reduced efficacy 0/16 (0%) 4/16 (25%)
Failed administration 2/16 (13%) 7/16 (44%)
Error rate 3.6% 20.1%
Mean total rescue time (start at entry of room until delivery of dose) 47.9 s 109.0 s
Median total rescue time (start at entry of room until delivery of dose) 35.0 s 96.0 s