Table 4.

Infusion and alarm frequencies by care area and across the whole hospital between the study dates January 1, 2000, to September 29, 2017 (6482 days).

Parameter Frequency in critical care areas, n Frequency in whole hospital, n Alarm frequency, %

All NICUa GICUb CCUc PICUd

Total infusion starts 187,441 30,527 45,756 110,254 904 1,600,832 N/Ae
Alarms per infusion, mean 4.50 3.71 4.36 1.33 8.61 1.39 N/A
Alarm types







Total alarms 467,437 113,277 199,482 146,694 7784 2,211,457 100

Flow error (ie, drip counter) 0 N/A 0 0 N/A 0 0

Air-in-line accumulation exceeded 6975 N/A 6700 275 N/A 27,583 1.25

Air-in-line single bubble exceeded 17,100 N/A 12,724 4376 N/A 108,701 4.92

Callback 134,754 22,396 68,037 42,740 1581 802,691 36.30

Door open while infusing 14 N/A 14 0 N/A 2710 0.12

Drive engage failure 10,850 5629 175 4624 422 13,807 0.62

End of infusion 20,762 4650 849 13,995 1268 23,903 1.08

Near end of infusion 107,236 10,695 49,517 45,668 1356 278,969 12.61

Occlusion (downstream) 136,148 62,944 49,444 21,314 2446 847,438 38.32

Occlusion (upstream) 14,602 N/A 11,710 2892 N/A 86,592 3.92

Syringe disengaged 15,400 6738 290 7915 457 15,288 0.69

End of syringe 3396 225 22 2895 254 3775 0.17
Infusions started from within the drug library (ie, detectable as critical infusions), % N/A N/A N/A N/A N/A 67.39 N/A

aNICU: neonatal intensive care unit.

bGICU: general intensive care unit.

cCCU: coronary care unit.

dPICU: pediatric intensive care unit.

eN/A: not applicable.