TABLE I.
Design characteristics of PET scanners used for comparison.
Parameters | UHR | HRRT [17] | Biograph Vision [19] |
---|---|---|---|
Scintillator materials | LYSO | LSO/LYSO | LSO |
Number of scintillators | 129,024 | 119,808 | 60,800 |
Scintillator size (mm3) | 1.12 × 1.12 × 12 | 2.1 × 2.1 × (10 + 10) | 3.2 × 3.2 × 20 |
Scintillator array dimension | 4 × 8 | 72 × 104 (× 2 layers) | 10 × 20 |
Photodetector | APD | PMT | SiPM |
Scintillator-to-photodetector ratio | 1 : 1 | 7488 : 140 | 200 : 128 |
Scintillators per ring | 896 | 576 (× 2 layers) | 760 |
Number of rings | 144 | 104 | 80 |
Axial length (mm) | 235 | 252 | 263 |
Ring diameter (mm) | 390 | 469 | 820 |
Transaxial FOV (mm) | 271 | 312 | 700 |
Packing fraction (%) | 56 | 71 | 91 |
Coincidence time window (ns) | 6 | 6 | 4.73 |
Energy window (keV) | 250–650 | 350–650 | 435–650 |
Dead time (ns) | 490 | 320 | 320 |