TABLE I.

Design characteristics of PET scanners used for comparison.

Parameters UHR HRRT [17] Biograph Vision [19]
Scintillator materials LYSO LSO/LYSO LSO
Number of scintillators 129,024 119,808 60,800
Scintillator size (mm3) 1.12 × 1.12 × 12 2.1 × 2.1 × (10 + 10) 3.2 × 3.2 × 20
Scintillator array dimension 4 × 8 72 × 104 (× 2 layers) 10 × 20
Photodetector APD PMT SiPM
Scintillator-to-photodetector ratio 1 : 1 7488 : 140 200 : 128
Scintillators per ring 896 576 (× 2 layers) 760
Number of rings 144 104 80
Axial length (mm) 235 252 263
Ring diameter (mm) 390 469 820
Transaxial FOV (mm) 271 312 700
Packing fraction (%) 56 71 91
Coincidence time window (ns) 6 6 4.73
Energy window (keV) 250–650 350–650 435–650
Dead time (ns) 490 320 320