TABLE 2.

Summary of results for LDBio Aspergillus ICT IgG-IgM test and routine serologya

Test % sensitivity (95% CI)
% specificity
(95% CI)
(healthy controls
[n = 148])c
Youden's
index
(all sera
[n = 302])d
DOR
(95% CI)
(all sera
[n = 302])
All CPA
(n = 154)
High
positive
(n = 11)b
Seronegative
(n = 10)b
LDBio ICT 91.6 (86.0, 95.4) 100 (71.5, 100) 90.0 (55.5, 99.8) 98.0 (94.2, 99.6) 0.896 524 (146, 1,879)
ImmunoCAP 80.5 (73.4, 86.5) 100 (71.5, 100) 0 (0, 30.9) NAe NA NA
McNemar's (P value) 0.007 NA
a

Results are reported against the cutoff of 40 mgA/liter recommended for use in United Kingdom laboratories at the time of publication.

b

ImmunoCAP results were used to define the “high positive” and “seronegative” groups.

c

Equivocal results (n = 2) were excluded from analysis.

d

Youden's index = sensitivity + specificity − 1.

e

NA, not applicable.