. 2019 Aug 19;40(1):33–39. doi: 10.3343/alm.2020.40.1.33

Table 2. Total precision of chemiluminescent immunoanalyzer-based interferon gamma releasing assay.

Samples Site Target value (IU/mL) Mean value (IU/mL) Repeatability Between-Run Between-Day Within-lab Reproducibility
SD % CV SD % CV SD % CV SD % CV SD % CV
1 1 90.21 86.49 4.20 4.86 4.67 5.40 3.73 4.31 5.80 6.70 5.46 6.36
2 85.12 4.91 5.77 3.37 3.96 2.70 3.18 5.01 5.88
2 1 34.30 37.14 2.34 6.30 1.86 5.01 1.39 3.75 2.53 6.82 2.74 7.55
2 35.33 2.18 6.18 1.55 4.39 1.37 3.89 2.34 6.62
3 1 2.47 2.38 0.13 5.60 0.10 4.11 0.05 2.31 0.13 5.43 0.17 7.24
2 2.25 0.13 5.79 0.10 4.49 0.08 3.76 0.14 6.40
4 1 1.05 1.07 0.07 7.00 0.06 5.69 0.05 4.48 0.08 7.80 0.08 7.88
2 1.02 0.05 4.99 0.04 3.7 0.05 4.45 0.06 6.25

Abbreviation: % CV, coefficient of variation expressed as percentage.