Table 2. Total precision of chemiluminescent immunoanalyzer-based interferon gamma releasing assay.
Samples | Site | Target value (IU/mL) | Mean value (IU/mL) | Repeatability | Between-Run | Between-Day | Within-lab | Reproducibility | |||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV | ||||
1 | 1 | 90.21 | 86.49 | 4.20 | 4.86 | 4.67 | 5.40 | 3.73 | 4.31 | 5.80 | 6.70 | 5.46 | 6.36 |
2 | 85.12 | 4.91 | 5.77 | 3.37 | 3.96 | 2.70 | 3.18 | 5.01 | 5.88 | ||||
2 | 1 | 34.30 | 37.14 | 2.34 | 6.30 | 1.86 | 5.01 | 1.39 | 3.75 | 2.53 | 6.82 | 2.74 | 7.55 |
2 | 35.33 | 2.18 | 6.18 | 1.55 | 4.39 | 1.37 | 3.89 | 2.34 | 6.62 | ||||
3 | 1 | 2.47 | 2.38 | 0.13 | 5.60 | 0.10 | 4.11 | 0.05 | 2.31 | 0.13 | 5.43 | 0.17 | 7.24 |
2 | 2.25 | 0.13 | 5.79 | 0.10 | 4.49 | 0.08 | 3.76 | 0.14 | 6.40 | ||||
4 | 1 | 1.05 | 1.07 | 0.07 | 7.00 | 0.06 | 5.69 | 0.05 | 4.48 | 0.08 | 7.80 | 0.08 | 7.88 |
2 | 1.02 | 0.05 | 4.99 | 0.04 | 3.7 | 0.05 | 4.45 | 0.06 | 6.25 |
Abbreviation: % CV, coefficient of variation expressed as percentage.