Table 3. Qualitative comparison of CLIA-IGRA and QFT-GIT assay results in health care workers and patients (N=341).

CLIA-IGRA QFT-GIT % Agreement (95% CI) Kappa coefficient
Positive (N=197) Indeterminate (N=1) Negative (N=143) Positive Negative Overall
Positive (N = 194) 194 0 0 98.48 (95.25–99.61) 97.95 (93.65–99.47) 99.12 (97.23–99.77) 0.98 (0.96–1.00)
Indeterminate (N = 1) 0 1 0
Negative (N = 146) 3 0 143

Abbreviations: CLIA-IGRA, chemiluminescent immunoanalyzer-based interferon gamma releasing assay; QFT-GIT, QuantiFERON-TB Gold In-Tube; CI, confidence interval.