Skip to main content
. 2019 Aug 23;10:1822. doi: 10.3389/fmicb.2019.01822

FIGURE 8.

FIGURE 8

Reproducibility of in-house indirect IgG ELISA. (A) Intra-assay variation (variation of replicates in one experiment) of IgG iELISA was performed with a panel of one from each low, medium, and high positive sample and three negative samples and five replicates within a plate. (B) Inter-assay-variation (variation (between experiments) performed on eight different days (days 1, 7, 10, 12, 17, 21, 30, and 40) in duplicates by the same operator. (C) Inter-operator variation (variation between experiments performed by different operators) and assay precision were conducted at different institutional laboratories and at FMD Centre, Gujarat, with a panel of seven positive and nine negative samples with different operators. Data interpretation was performed with Kruskal–Wallis and one-way ANOVA.)