Table 1:

Imaging parameters for ungated QISS and CE-MRA sequences

Parameter Ungated QISS-MRA CE-MRA
Imaging mode 2D 3D
FLASH TR/TE (ms) 15.0/4.7 3.09/1.2
QISS sequence TR (ms) 1100.8
Acquisition matrix (Px) 384 × 384 512 × 512
Acquisition pixel (mm2) 0.5 × 0.5 0.6 × 0.6
In-plane interpolation On On
Slice thickness (mm) 2.0 1.0
No. of slices 128 80
Slice distance factor (%) −33 20
No. of averages 1 1
Receiver bandwidth (Hz/Px) 303 540
Flip angle 30° 30°
Slice orientation Tilted transversal to coronal (45° tilt) Coronal
K-space trajectory Radial Cartesian
No. of radial projections 204
No. of shots per slice 3
Phase oversampling (%) 0 40
Filter Distortion correction (2D); prescan normalizer Distortion correction (3D); prescan normalizer
B0 shim mode Heart Tune-up
Asymmetric echo Off On
RF pulse type Normal Normal
Gradient mode Fast Fast
RF spoiler On On
iPAT modus (acceleration factor/No. of reference lines) 2/24
Partial Fourier (phase and slice) 6th/8th
Venous saturation slab thickness (mm) 100
Distance between venous saturation and imaging slab (mm) 10
TI (ms) 530
Acquisition time (min:sec) 7:03 0:20

Note:—iPAT indicates integrated parallel imaging technique; TI, time from in-plane and venous saturation to the acquisition of central k-space (ky = 0); Px, pixel; –, sequence parameter is not available; TR, repetition time; TE, echo time; RF, radiofrequency; Hz, Hertz.