Table 2.
Validation strains | No. of strains | Identification by VITEK MS | Confidence values (%) | Percentage |
---|---|---|---|---|
VIM-type MBL-PA | 4 | VIM-type MBL-PA | 99.9 | 16.0 |
12 | VIM-type MBL-PA | 90.0–99.9 | 48.0 | |
9 | VIM-type MBL-PA | 80.0–90.0 | 36.0 | |
Subtotal | 25 | VIM-type MBL-PA accuracy rate | 100.0 | |
SPM-type MBL-PA | 5 | SPM-type MBL-PA | 99.9 | 20.0 |
8 | SPM-type MBL-PA | 90.0–99.9 | 32.0 | |
8 | SPM-type MBL-PA | 80.0–90.0 | 32.0 | |
4 | VIM-type MBL-PA | 90.0–99.9 | 16.0 | |
Subtotal | 25 | SPM-type MBL-PA accuracy rate | 84.0 | |
Total | 50 | Total accuracy rate | 92.0 |
Note: The validation results showed that all the VIM-type MBL-PA isolates were correctly identified, just four SPM-type MBL-PA isolates were incorrectly identified as VIM-type MBL-PA isolates which got conflicting results in repeat testing by VITEK MS plus system.