. 2019 Sep 6;12:2781–2788. doi: 10.2147/IDR.S211984

Table 2.

Results of validation of VITEK MS plus system in identification of VIM- and SPM-type MBL-producing P. aeruginosa isolates using the two SuperSpectra

Validation strains No. of strains Identification by VITEK MS Confidence values (%) Percentage
VIM-type MBL-PA 4 VIM-type MBL-PA 99.9 16.0
12 VIM-type MBL-PA 90.0–99.9 48.0
9 VIM-type MBL-PA 80.0–90.0 36.0
Subtotal 25 VIM-type MBL-PA accuracy rate 100.0
SPM-type MBL-PA 5 SPM-type MBL-PA 99.9 20.0
8 SPM-type MBL-PA 90.0–99.9 32.0
8 SPM-type MBL-PA 80.0–90.0 32.0
4 VIM-type MBL-PA 90.0–99.9 16.0
Subtotal 25 SPM-type MBL-PA accuracy rate 84.0
Total 50 Total accuracy rate 92.0

Note: The validation results showed that all the VIM-type MBL-PA isolates were correctly identified, just four SPM-type MBL-PA isolates were incorrectly identified as VIM-type MBL-PA isolates which got conflicting results in repeat testing by VITEK MS plus system.