. 2019 Aug 29;11(4):384–391. doi: 10.5114/jcb.2019.86973

Table 1.

Results of the doses (Gy) from the ABV calculation algorithm and measured doses from the TLD and MOSFET dosimeters at the various points in the in-house phantom

Point ABV MOSFETUNC TLDCOR MOSFETUNC/TLDCOR MOSFETCOR/ABV TLDCOR/ABV
1 6.30 6.46 6.51 0.99 0.99 1.03
2 6.15 6.52 6.42 1.02 1.03 1.04
3 1.73 1.82 1.77 1.03 1.02 1.02
4 8.35 9.19 8.84 1.04 1.07 1.06
5 8.01 8.55 8.38 1.02 1.04 1.05
6 1.72 1.81 1.79 1.01 1.02 1.04
7 9.95 10.58 10.02 1.06 1.03 1.01
8 8.69 9.11 8.70 1.05 1.02 1.00
Mean 6.36 6.75 6.55 1.03 1.03 1.03
SD 3.12 3.34 3.18 0.02 0.02 0.02

UNC – readings were uncorrected for beam quality differences, COR – readings were corrected with the energy response factor for 192Ir