Skip to main content
. 1974 Jul-Aug;78A(4):465–476. doi: 10.6028/jres.078A.030

Table 13.

Chamber correction kstd required to bring exposure-rate measurements for each chamber into agreement with the weighted mean exposure rate X˙¯.

Chamber 60Co 137Cs
1 0.9970 0.9962
10 .9997 .9997
30 1.0005 1.0022
50–1 1.0002 0.9996
50–2 1.0003 .9996
50–3 0.9998 .9995