Table 1.
Score Stage |
Optotype Gray RGB Value |
Background Gray RGB Value |
Actual Peek CS Log10 (CS) |
Standard Error |
Pelli-Robson Log10 (CS) |
Contrast |
1 | 0 | 255 | 0.006 | 0.000007 | 0.00 | Highest |
2 | 164 | 255 | 0.20 | 0.00040 | 0.15 | E |
3 | 197 | 254 | 0.35 | 0.000625 | 0.30 | |
4 | 216 | 255 | 0.50 | 0.00026 | 0.45 | |
5 | 229 | 255 | 0.65 | 0.001225 | 0.60 | |
6 | 237 | 255 | 0.78 | 0.00306 | 0.75 | |
7 | 242 | 254 | 0.96 | 0.00208 | 0.90 | |
8 | 246 | 255 | 1.06 | 0.03935 | 1.05 | |
9 | 248 | 255 | 1.16 | 0.0017 | 1.20 | |
10 | 250 | 254 | 1.40 | 0.01185 | 1.35 | |
11 | 250 | 253 | 1.53 | 0.0184 | 1.50 | |
12 | 252 | 255 | 1.67 | 0.00475 | 1.65 | |
13 | 246 | 248 | 1.71 | 0.00875 | 1.80 | |
14 | 252 | 254 | 1.94 | 0.0034 | 1.95 | |
15 | 254 | 255 | 2.00 | 0.0076 | 2.10 | E |
16 | N/A | N/A | N/A | N/A | 2.25 | Lowest |
N/A, not applicable; RGB, red, green, blue.
Designed for a Sony Xperia Z3 for each Score Stage and the corresponding actual CS measured by a photometer in a darkened laboratory setting. Owing to the finite combinations of optotype and background grays, it is not possible to have an exact alignment between the contrast sensitivities tested with PeekCS and the Pelli-Robson charts. However, the difference between the two tests is never more than two-thirds of the difference between score stages (i.e., 0.10 log units).