Skip to main content
. 2019 Aug 29;17:e00131. doi: 10.1016/j.plabm.2019.e00131

Fig. 1.

Fig. 1

Effect of Scavenger ALP on E2 and uE3 Measurement. Three different patient specimens identified by prior IND error flags for uE3 and/or E2 were tested on UniCel DxI 800 assays for E2 (A) and uE3 (B) with (+) and without (-) incubation with 500 μg/mL Scavenger ALP. Note that as a competitive immunoassay, increased RLUs correlate to decreased apparent analyte concentrations.