Skip to main content
. 2019 Aug 29;17:e00131. doi: 10.1016/j.plabm.2019.e00131

Table 2.

IND Error Flags with E2 Results in Six Patients.

Pt. A Pt. B Pt. C Pt. D Pt. E Pt. F
Age, yrs 46 29 42 49 57 55
Ordered Test E2 E2 E2 E2 E2 E2
E2 Undiluted IND IND IND IND IND IND
E2 1:2 in Sample DilA IND IND <20 pg/mL <20 pg/mL IND IND
Mass Spec Detection of E2 N/A 29.6 pg/mL 17.6 pg/mL N/A N/A 59.2 pg/mL
Interference Suspected Yes Yes No No Yes Yes