Skip to main content
. 2019 Aug 29;17:e00131. doi: 10.1016/j.plabm.2019.e00131

Table 3.

IND Error Flags with uE3 Results in Seven Patients.

Pt. G Pt. H Pt. I Pt. J Pt. K Pt. L Pt. M
Age, yrs 21 24 34 47 29 18 19
Ordered Test Quad Screen Quad Screen Quad Screen uE3
only
Quad Screen Quad Screen Quad Screen
uE3 Undiluted IND IND IND IND IND IND IND
uE3 1:2 in Sample DilA IND IND <0.02 ng/mL IND 0.11 ng/mL <0.02 ng/mL <0.02 ng/mL
Mass Spec Detection of uE3 Present Present Absent Absent N/A N/A N/A
Interference Suspected Yes Yes No Yes No No No

N/A, Not Applicable, Insufficient specimen for additional testing.