. 2019 Feb 9;48(5):20180301. doi: 10.1259/dmfr.20180301

Table 1.

Specifications of HIDXDs

Device
(Manufacturer)
Fixed tube voltage of HIDXD (kVp) Fixed tube current of HIDXD (mA) Manufacturer specified lead equivalence of backscatter shield (mm Pb) Diameter of backscatter shield (cm) Manufacturer specified total filtration of HIDXD (mm Al equivalent) Measured half-value layer of HIDXD (mm Al equivalent) Measured Air Kerma per current-time product of HIDXD (mGy/mAs)
BIOX-IPX 0
(DigiMed Co., Ltd., Seoul, Korea)
60 3 Not stated 15.5 2.8 2.87 0.887
Nomad Pro2
(Aribex, Inc., Orem, UT)
60 2.5 Not stated 15.5 ≥1.5 2.22 1.473
SAF-3000
(Safari Dental Inc., Boisbriand, Québec)
60 2 0.35 @ 70 kVp 14.8 2.3 2.40 1.264
Xray2Go
(Osstem Implant Co., Ltd., Geumcheon-gu, Korea)
60 2 Not stated 15.8 1.6 2.76 0.830
Zen-PX2
(Genoray Co., Ltd., Seongnam, Korea)
60 2 Not stated 15.0 1.8 2.76 0.720

HIDXD, hand-held intraoral dental X-ray device.

Summary of select device specifications and X-ray beam quality for the five HIDXDs evaluated for stray radiation. Backscatter shields were in place for all devices. The source-to-skin distance for each device was 20 cm with the removable cones in place. A timer setting of 1.0 sec was used for the measured air kerma output for all devices. For each device, the timer setting is the only adjustable loading factor, thus differences in output for a given device are solely a result of changes to selected exposure time.