Table 2.

Quantitative summary of measured stray radiation air kerma values

Device Current–time product
(mAs)
Median air kerma (nGy) Mean air kerma (nGy) Max air kerma (nGy) Standard deviation Interquartile range
(Q3–Q1)
BIOX-IPX 0 3.0 46 138 438 0.14 234
Nomad Pro2 2.5 124 289 958 0.28 447
SAF-3000 2.0 89 213 666 0.20 340
Xray2Go 2.0 46 138 438 0.14 234
Zen-PX2 2.0 138 182 459 0.14 256

Summary of key quantitative metrics for measured stray radiation air kerma values at all 63 spatial locations within the Significant Zone of Occupancy (SZO). A timer setting of 1.0 sec was used for all device measurements, implying the X-ray output is not constant across the devices. Normalization is required to directly compare the measured stray radiation between devices.