Quantitative summary of normalized stray radiation air kerma values (SNorm)

Device Median air kerma (nGy) Mean air kerma (nGy) Max air kerma
(nGy)
Median air kerma typically required for D-speed film (1.6 mGy air kerma at end of dental cone)
BIOX-IPX 0 28 83 264
Nomad Pro2 54 126 416
SAF-3000 56 135 421
Xray2Go 44 133 423
Zen-PX2 154 203 510
Mean 67 136 407
Median air kerma typically required for digital receptor (0.8 mGy air kerma at end of dental cone)
BIOX-IPX 0 14 41 132
Nomad Pro2 27 63 208
SAF-3000 28 67 211
Xray2Go 22 66 211
Zen-PX2 77 101 255
Mean 34 68 203

Measured stray radiation air kerma values as normalized to the median air kerma values typically required to produce a diagnostically acceptable image on a digital receptor (0.8 mGy) and D-Speed film (1.6 mGy).